SE2 > MS1
ID SE2_MS1
Title LC-FT-ICR-MS ESI positive method 1
Instrument Agilent1100 HPLC (Agilent), LTQ-FT (Thermo Fisher Scientific)
Instrument Type LC-FTICR-MS
Ionization Method ESI
Ion Mode Positive
Description Harvested sample is frozen by liquid N2 and resulting powder (100mg) are solved in 300uL 80% methanol solution. 20uL sample is injected into HPLC after 0.2um membrane filter treatment. HPLC conditions: Agilent 1100 series (Agilent), Column: TSKgel-100V (4.6 x 250 mm, 5 micrometer; TOSOH), Solvent: A; 0.1% formic acid aq. B; ACN (addition 0.1% formic acid fc.), Gradient: (B);3 to 30% (0.0 to 25.0 min), 30 to 90% (25.0 to 40.0 min), 90% (40.0 to 45.0 min), 95% (45.1 to 50.0 min), 3% (50.1 to 57.0 min), Column temp.: 30 degree C, Flow rate=0.5mL/min, PDA: 200-650 nm (2 nm step). FT-ICR-MS conditions: Filter 1: FTMS + c norm !corona !pi res=50000 o(200.0-1500.0); 2: ITMS + c norm !corona !pi Dep MS/MS Most intense ion from (1).;3: ITMS + c norm o(200.0-1500.0)., Rejected mass=266.0000;294.0000;391.0000.
Comment_of_details -