Title mQTL analysis
Instrument UPLC-QTOF Premier (Waters) and UPLC-TQS (Waters)
Instrument Type UPLC-QTOF-MS
Ionization Method ESI
Ion Mode N.A.
Description The metabolites were analyzed using a liquid chromatography quadrupole time of flight mass spectrometer and tandem quadruple mass spectrometer system (UPLC-QTOF Premier and UPLC-TQS, Waters Co., MA, USA) (Matsuda et al., 2009; Sawada et al., 2009a).
After cutting off the low intensity data of MS2Ts, 702 MS2Ts were used for a selected reaction monitoring (SRM) assay of UPLC-TQS. To determine the optimal collision energy for MS2Ts, collision induced dissociation fragmentation analyses at six energy steps (10–60 eV) were conducted using the same seed extracts. As a result, 4212 SRMs were used for analysis. The SRM condition that gave the highest signal intensity among the six energy steps was defined as the optimal SRM. The optimal SRM conditions were further selected based on the following criterion: the relative standard
deviation of UPLC-TQS peak area values must be less than 10% (analytical replicates = 3). Finally, 342 SRM conditions for MS2Ts were successfully assigned to the seed extract of L. japonicus. Based on the natural variations in the metabolite accumulation patterns among the accessions (MG-20 and B-129), 88 SRM conditions were used for mQTL analysis.


<mQTL analysis>

Metabolic profiling data of 129 RILs were processed as follows: the peak area values of 88 SRMs were divided by the peak area values of the internal standards, which were lidocaine and camphor sulfonic acid for the positive and negative ion modes, respectively. The divided values were then converted into z-scores of binary logarithms after missing values were replaced with 0.1. The genotype data of each RIL were obtained from KDRI (http://www.kazusa.or.jp/lotus/). Using processed metabolic profiling data and gene marker data of RILs, mQTL analysis was carried out using R/qtl (Broman et al., 2003) (Supplementary Data).
Comment_of_details -